NOAH N Electrochemistry Report FINAL2

 

Stage 1 Chemistry

Electrolysis Practical Investigation

 

 

 

 

 

 

 

 

 

Name: Noah Nishihara

Home Group: E05

Report of experiment on:

 The Mass Difference of Electrodes inthe Electrolysis of Copper

Date of Experiment: 26 May 2017

Work Finalised: 2 June 2017

 


 

 

 

 

 

 

 

Table of Contents

Aim.. 4

Outline of Procedure. 4

Results. 9

Calculations. 11

Discussion. 14

Conclusion. 17

 


 

Aim

To investigate the effect of electrolyte concentration on the process of copper refining, this experiment will determine the anode mass loss during ten minutes of copper electrolysis.It will also attempt to confirm that the theoretical amount of mass lost from the anode equals the actual amount of mass lost from the anode copper electrode.

Outline of Procedure

Apparatus

·         100 mL and 250 mL size beakers

·         100 mL measuring cylinders

·         Copper electrodes of assorted sizes

·         Alligator clips to attach electrodes onto the side of the beaker

·         Two insulated wire pieces

·         DC capable power supply (regulated)

·         Aqueous copper(II) sulphate, 0.5 M

·         Paper towels

·         Steel wool

·         Stopwatch

·         Ruler

·         Incubator

·         Top loading balance (max 450 g)

·         Blu-tack

·         Distilled water (in containers)

 

 

 

 

 

Diagram:Anode copper mass loss before and after the electrolysis; comparison of the practical results and theoretical results and a change of concentration of electrolyte

 

 

 

 

Procedure Description:

This experiment was carried out twice with the same concentration of 0.5 M copper sulphate and then repeated twice with 0.2 M copper sulphate for a total of four measurements.

1.It was ensuredthat the copper electrodes were clean by rubbing them with steel wool. This resulted in shiny electrodes without the possibility of contamination or moisture which could potentially affect data collection and measurements.

2.The mass of the two labelled (+and - ) copper electrodes were measured separately with a top loading balancewith the lid covering it andthe results were recorded.

3.90 mL of 0.5 mol / L copper (II) sulphate aqueous solution was addedinto a measuring cylinder,

4.This was poured into a 100 mL beaker to assemble the circuit shown in the figure above.

5.The electrodes were clamped to the rim of the beaker using crocodile clips ensuringthat the electrodes did not touch. The distance between electrodes was kept as constant as possible.

6.The clips were not touching the electrolyte. The positive anode copper was connected with the positive end of the power supply. The negative cathode was connected to the negative end of the power supply.

7.The area of the two copper plates immersed in the copper (II) sulphate aqueous solution waskept as similar as possible.

8.The power supply was set to 2.0 V to prevent water from reacting and producing hydrogen ions H+ and hydroxide ions (OH).

9.The current intensitywas set to 3.1 A (or equivalent maximum value)

10.Current was allowed to flow for 10 minutes (600 seconds) by pressing the on button and starting the stopwatch simultaneously.

11.The currentreading was continuously monitored and recorded along with the time on the stopwatch.

12.Movement of the electrodes was avoided.

13.The copper electrodes were removed, rinsedcarefully withdistilled water and left to evaporate on blu-tack (for the cathode mass gain measurement). The anode was wiped and the copper oxide was not included in the mass measurements.

14.The mass of each copper plate was weighed.

15.The process was repeated another three times in totalwith 0.2 mol/L of copper (II) sulphate and different electrodes which were weighed again.

 

Principles

The measuring principle utilised here is integer counting. The data is collected directly as values from measuring devices. The top loading balance used had a resolution of 1 to 0.001 g and could weigh up to 450 grams. Calibration of the device was not carried out as the digital readout was deemed correct before use.

Basic Formulas Used

Concentration Calculations: C1V1 = C2V2

To find Charge:

Formulas used in analysis

Number of moles:

Note:-1.602×10-19C× 6.022×1023/mol -9.65×104C/molSince the quantity of electricity Q of one electron is constant at -1.602 × 10 -19 C, in electrolysis, the amount of changing substances is proportional to the quantity Q of electricity passing through.

Which is

Hypothesis:If the concentration of copper (II) sulphate is raised, then the efficiency of the reaction will also rise in the electrolytic refinement of copper.

Variables

Independent Variable: Concentration of copper (II) sulphate electrolyte

How the Independent Variable is changed: Alternate the electrolyte conditions from 0.5M to 0.2M(Copper (II) sulphate concentration)

Dependent Variable: The amount of the anode copper that is lost and from that, the efficiency can be derived (more mass loss may mean higher efficiency)

How the Dependent Variable is Measured: Mass measuring top loading balance,

Other factors held Constant in the Experiment:

  • Type of electrodes –Copper anode and cathode – anode quality
  • Time taken to complete reaction – 10 minutes(stopwatch for 10 minutes of time)
  • Volume of solutions used in the electrolysis reaction
  • Surface area of electrode dipping into the electrolyte solutions – measured to 5 cm
  • Maintain voltage with regulated power supply – 0.15V-0.30V (industrial) here it is 2.0 V
  • Current of the electrolysis circuit – 10-20 mA/cm2(industrial) 0.02 A if lengthy reaction – cathode current density - rate at which the anode is dissolved and the rate at which the cathode is plated
  • Attempt to maintain constant temperature – or take into consideration its fluctuations – recommended industrially to be 60 degrees Celsius (22 degrees Celsius)
  • Distance between electrodes – for the 1st and 3rd reactions the distance was less

 

 

Results

Table 1 Main Results

Order of Attempt

Type of Electrode

Concentration (Mol/L)

Mass Before Electrolysis (grams)

Mass After Electrolysis (grams)

Difference in Mass (grams)

1

Anode

0.5

0.978

0.897

0.081

1

Cathode

0.5

0.88

0.944

0.064

 

 

2

Anode

0.5

1.617

1.574

0.043

2

Cathode

0.5

0.94

0.981

0.041

 

 

3

Anode

0.1

8.809

8.777

0.032

3

Cathode

0.1

8.936

8.972

0.036

 

 

4

Anode

0.1

1.13

1.107

0.023

4

Cathode

0.1

0.862

0.886

0.024

 

Table 2 Current Changes of the Reactions

1st

Recording

2nd

Recording

3rd

Recording

4th

Recording

Time (s)

Current (A)

Time (s)

Current (A)

Time (s)

Current (A)

Time (s)

Current (A)

0

0.325

0

0.247

0

0.157

0

0.117

50

0.326

60

0.248

40

0.158

3

0.118

80

0.327

195

0.249

95

0.159

5

0.119

100

0.328

480

0.25

138

0.16

25

0.118

120

0.328

600

0.25

195

0.161

220

0.119

230

0.329

250

0.162

270

0.119

318

0.33

320

0.163

390

0.119

445

0.331

430

0.164

490

0.12

588

0.332

520

0.165

600

0.12

600

0.165

The distance between electrodes was shortened for the 1st and 3rd recording of data which may have contributed to the increased change in current. A constant distance may have different results to the above.Changes in current were greater the less distance between the electrodes.

Using the 5th Equation, the percent error formula, the results of the experiment (anode mass losses) were compared with the theoretical calculations of the mass lost at the anode. The second attempt was shown to have the greatest error.

The results are shown in the table below:

Table 3 Percent Error of Reactions

Error with Lowest Current (%)

Error with Highest Current (%)

1

0.3115

2.439

2

11.885

12.955

3

3.226

1.84

4

0.433

2.954

Calculations

Following on from the Percent Error Formula calculations which indicate the ratio of the observed and true values, the 6th equation is used to find the percent uncertainty. For the first attempt at the lowest current, it is 4.867%. This is given by:

The experimental result is compared with the theoretical result in % error, while % uncertainty is found to give a better idea of how well a value can be determined. This uncertainty value also reveals how precise the experiment was. The values for uncertainty are shown below:

Table 4 Uncertainty in Measurements

Uncertainty with Lowest Current (%)

Uncertainty with Highest Current (%)

1

4.867

38.109

2

276.395

301.279

3

100.813

0.575

4

18.826

128.435

The values somewhat correlate with the values for percent error.

The statistical standard deviation was found for the mass loss using both theoretical values of low and high current and the experimental value.

Table 5 Standard deviation of Mass loss

Mean (g)

Standard deviation

1

0.0646

±0.0008718

2

0.0471

±0.003533

3

0.0319

±0.0008083

4

0.0233

±0.0003843

This calculation falsely leads to the belief that the experiment was not too spread far out. This was not the case, with an alternative interpretation possible of the second attempt being the most error-ridden and the first and third attempts at around the same variance. Only the fourth attempt has satisfactory accuracy. It should be noted however, that the electronic top-loading balance used only measured to the third decimal place.

Concentration:

Moles of copper atoms:

Since copper (II) ionsare divalent cations, two electrons are necessary for one copper (II) ion to become a copper atom.

Moles of electrons:

10 minutes are 600 seconds

 

 

 

 

 

 

The values are summarised in the table below:

Table 6 Moles of electrons comparison

Moles Of Electrons – From Experiment Data Of Electrode Mass Difference (Mol)

Moles Of Electrons – Theoretical Values (Lowest Current) (Mol)

Moles Of Electrons – Theoretical Values (Highest Current) (Mol)

 

Calculations of: Theoretical mass lost/gained of Copper

Moles of Copper

Mass of Copper:

The amount of charge gives the mass of copper lost or gained in the calculations using equation 4. These values were used for the percent error calculations and revealed the second attempt to be the most inaccurate, while the first mass measurement of the anode which gave 0.081 grams was at a 26% deviation.

This is summarised in the table below:

Table 7 Mass change-accepted values

Mass at Low Current (g)

Mass at High Current (g)

1

0.0642

0.0656

2

0.0488

0.0494

3

0.031

0.0326

4

0.0231

0.0237

Discussion

The Reactionduring electrolysis:

A transfer of copper metal from the anode to the cathode is the objective.

Each copper atom loses two electrons (oxidation) to become copper ionsat the anode. Thecopper cations move to the cathode, where two electrons are gained (reduction) to become copper atoms that are deposited on the cathode.The sulphate negatively charged ion moves to the anode but does not reactas it is stable and the concentration of ions is constant as the reaction of copper occurs at both ends. The reactivity and concentration of ions determines this.

The theoretical values attained using Equation 2 were lower for the first attempt, and greater for the second and fourth attempts compared to experimental values. Only the third attempt fitted the theoretical values. The first resultwas due to errors in the measuring of mass. The cleansing process may have affected the values for others. The current reading could have been lower than the actual amount, causing the data to be lower. Contrastingly, the low error of the fourth attempt indicated a close fit to the theoretical mass values, further proving the misleading nature of the values of moles of electrons calculated with Eq.2.

The validity of the 0.5M solution data was questionable from the error calculations. However, the pattern without taking errors into account was: the closer distance between electrodes gave greater increase in current.The first and third reaction was carried out with a distance between the electrodes of close to one centimetre. The current increased by close to 0.01 Ampere for this distance. (The distance was difficult to control due to the nature of clipping electrodes to the beaker.)The change in mass was greater for higher values of current as shownwith the theoretical calculations. Thus, increases in current did not determine the efficiency of the reactions. The 4th reaction with lower current gave 23 % to 29 % less efficiency than the 3rd reaction under similar conditions.This inefficiency value was predictedto increaseproportionally with greater current values,for instance, in the first reaction with 36 % to 47 % less efficiency.

The main findings of this experiment were mass loss values found to decrease with each experiment. The reactions resulted in different mass loss values of the anode. There was a correlation between the concentration and the mass loss, due to differing current. Higher concentrations had greater mass loss due to a raised current value.Lower concentrationshad less mass loss due to a lowered current value.

In the graphs of current change by time, it was observed that for the first reaction, the current increased by 0.003 A in the first two minutes, but then took over five minutes to increase by another 0.003A. Similarly, the second attempt saw more rapid increase of current in the first few minutes as opposed to later, when the value fluctuated until the time limit was reached. The third round had the greatest increase of current, with an increase of 0.005 A in four minutes and 0.003 A in another four minutes. The fourth reaction had the current fluctuating from 0.118 A to 0.119 A from 3:40 to 4:30. The increase overall was 0.003A. The amount of current change was reduced by longer distances between electrodes. Current change always rose steeply and then became stable.

Reaction conditions including voltage can be lowered to be more effective in returning metal ions to the original metal form. The experiment outlined here was performed under time constraints. Current was not set to allow the voltage to fluctuate due to lower performance. However, a controlled lower voltage over a longer period may be an improvement over the current experiment design.

The top-loading balance required a lid to be placed on top. It was uncalibrated due to lack of knowledge.

Slight differences in the surface area exposed to the electrolyte were unavoidable due to the different shapes of the electrodes. An attempt was made to use like-sized electrodes together and ensure that the length submerged was no more than 5 cm.

The drying of the cathode was done carefully with evaporation attempted and no contact to it by any material. The meniscus was found at eye level for the measuring and distribution of copper sulphate solution. Measuring cylinders of 100 mL capacity have less obvious formations of a meniscus, however, it was ensured that the bottom of the meniscus was precisely on the calibration mark.

Although the experimental values for the balance were to three decimal places, the theoretical mass values were to four decimal places. A balance with greater accuracy and less maximum measurable mass would have increased the precision of the data analysis.

Using the percent error formula, the second attempt was found to have a lower accuracy than the other attempts due to balance fluctuations. This may have been caused by a failure in the operation of the electrolysis. The other measurements of anode mass loss were within five percent of the theoretical values.

Systematic errors affected the accuracy of the practical in the form of the use of non-homogenous solutions. The result of this may be different current values and subsequently changes to the electrode mass loss, and would lower the accuracy of the collected data. To eliminate this error, the higher concentration could be raised to 1.0M and compared with a lower solution of 0.5M.

The accuracy of experimental values in this report could be increased if measuring instruments are calibrated or solutions are standardised. Different apparatus could be used.By repeating the experiment, the reliability can be increased. The validity of this experiment was high, as the overall data proves what was hypothesised, though with some measurement errors.

To apply the results contained within the report, further studies will be necessary. Distance between electrodes should be minimised for maximum increase in current. Higher concentrations of copper(II) sulphate work well because of the current increase.

Conclusion

High values of current with 0.5 M copper(II) sulphate saw greater efficiency than low current values with copper(II) sulphate of lower concentrations. The calculations of the theoretical mass loss of the anode reflected the actual mass loss for the 0.2 M copper(II) sulphate experiments only.

Comments

Popular posts from this blog

Vegetable People Practical

Single Text Essay Structure

COMPLETED VERSION Charles Perkins and the 1965 Australian Freedom Ride